  • 图片30.jpg
  • 图片31.jpg

RZSCF-3 Supercritical Water Reaction Equipment

It is the high tech company which is committed to the super critical fluid technology development and equipment manufacturing.

Share to

Product Description

  • 产品描述
  • Serial number

    Main configuration

    Parameter

    1

    Extraction kettle

    Volume: customized according to customer requirements, generally 100ml-1000l
    Working pressure: 35MPa-100MPa adjustable
    Working temperature: room temperature -85℃ adjustable
    3 extraction kettles recommended for production equipment

    2

    Separating kettle

    Volume: According to the volume of the extraction kettle, generally 50ml-800l
    Working pressure: 10MPa-30MPa adjustable
    Working temperature: room temperature -85℃ adjustable
    All equipment recommends 3 separation kettles

    3

    Purification system

    Working pressure: 16MPa

    4

    Storage tank

    Volume: According to the overall plan configuration, generally 10l-6000l
    Working pressure: 9.9MPa

    5

    Pump

    CO2 high pressure pump and carrier pump, adjustable flow and pressure
    Production equipment equipped with 2 high-pressure pumps

    6

    Heat exchange system

    Working pressure: 35MPa-100MPa
    Working temperature: adjustable from room temperature to 90℃

    7

    Refrigerating system

    5℃-10℃

    8

    Security system

    The high pressure pump and the inlet and outlet of all containers are equipped with overpressure protection device, the safety interlock device is connected with the PLC control system and the safety valve to ensure the safety of the system

    9

    Recycling system

    Configure corresponding recovery compressors according to the size of the equipment to recover excess CO2 and reduce losses

    10

    PLC control system

    All equipment can be equipped with PLC control system to automatically control extraction and separation pressure, flow, temperature and other parameters

    11

    User-supplied

    1. CO2 gas source (food grade>99.9%)

    2. Water source (deionized water)

    3. Refrigerant (ethylene glycol)

    4. Cooling water tower (configured according to equipment size)

    5. Steam (industrial equipment)

    6. Power supply (according to local requirements)

     

     